DIMATEC's DIMATOC 2100 combines the highest precision and flexibility for TOC analysis of wastewater and pure water, analysis of samples with different concentrations, use of different catalysts, etc. TOC measurements are based on the principle of thermo-catalytic oxidation and downstream NDIR detection. the DIMATOC 2100 features two completely independent measurement channels and a carry-over-free sample feed system, which makes it easy for even samples that contain particles to be analysed. samples containing particles can be easily mastered. all relevant components of the DIMATOC 2100 are easy to use, and the modular system design allows the instrument to be customised according to customer requirements and subsequently retrofitted and upgraded.
Product features
Two independent measuring channels: ensures flexibility and accuracy
DIN compliant: conforms to DIN EN 1484 and DIN EN 15936, amongst other relevant standards.
Multi-parameter measurement: TC, TIC and TNb can be measured simultaneously.
Excellent particle handling: even samples containing particles can be handled with ease
High sample throughput, short measurement times: increased efficiency
Low maintenance requirements: reduces operating costs
High durability and easy accessibility: ensures long-term stable operation of the device
Flexible, fast and versatile: the true 2-channel design offers versatile possibilities, e.g. differential methods for TOC measurement, use of different catalysts, applications with different measuring ranges, etc.
Simple to operate: the device is designed to be practical, functional and as simple as possible, making it easy to operate.
Cost-effective and future-proof: the modular design allows the equipment to be customised, upgraded and expanded as requirements change
Excellent service support: DIMATEC offers personalised advice and professional technical customer service.
Technical parameters
Measured parameters: TC, TIC, TOC=TC-TIC, DOC, POC, NPOC, optional TNb
Measurement method: Thermal catalytic oxidation up to 900°C, full dual channel NDIR detection
DIN compliance: TOC according to DIN EN 1484 and DIN EN 15936; TNb according to DIN EN 12260 (additional module DIMA-N required)
Sample processing: fully automated, including acidification, degassing, homogenisation
Particle handling capacity: up to 0.5 mm
Sample positions: 60 or 78 sample positions, 15-40 ml each (stirring/shaking possible) or individually customised
Measuring range: 0-20,000ppm C
Control: PC with DIMAQS® software
Device dimensions: approx. 60x66x76cm (including sample provider)