The TC125 temperature transmitters convert temperature-dependent resistors (Pt100 / Pt1000) and thermocouples (J, K, S) into standard signals (0/4..20 mA or 0/2..10 V). Applications with signals from the Ex area or requirements with higher interference immunity require galvanic isolation. This is especially true for grounded temperature sensors. The TC125 has a 3-way galvanic isolation between input, output and auxiliary voltage. For PT100, Pt1000 and thermocouples, 8 preconfigured measuring ranges each are available, which can be adapted with rotary potentiometers with regard to zero point and span. When using thermocouples, internal temperature compensation can be activated or deactivated by measuring the terminal temperature. Due to an intrinsically safe input, the use in applications in the EX area is possible.
Attributes of Temperature Transmitter TC125 - TC125M-EX-00
May differ from supplier data
Housing / Design / Mechanical data / Materials
Housing width 12.5 mm
Housing height 108 mm
Housing depth 114 mm
Sensor weight, approx. 120 g
Housing material PA6.6
Basic data
Accuracy (%) 0.2 %
Linearity 0.05 %
Output signal range (in V) 2 to 10 V
Output signal range (mA) 4 to 20 mA
Input quantity Pt100 Pt1000 Type J thermocouple Show all Type K thermocouple Type S thermocouple
Potentially explosive area Zone 2
Product environmental influences (protection class/temperature)
Storage temperature -20 to 80 °C
Operating temperature range 0 to 50 °C
Protection class IP20
More functions and options
More functions/options without option