SMARTPAT PH 8150
Item number: PH 8150 Applications Chemical industry
Price on request
Delivery time:Delivery time on request
* Current loop powered 2-wire sensor with integrated transmitter technology. * Offline calibration function. * Installation in hazardous area, zone 0.
Attributes of SMARTPAT PH 8150
May differ from supplier data
Housing / Design / Mechanical data / Materials
Standard shaft length/installation length 120 mm
Shaft diameter 12 mm
Basic data
Minimum conductivity > 150 µS/cm
Measurement range 0 to 14 pH
Measurement possibilities pH value
Diaphragm Open diaphragm
Sensor body Glass electrodes
Measured medium Wastewater
Product environmental influences (protection class/temperature)
Process pressure, max. 12 to 580 bar
Operating temperature range 0 °C
Approvals / Certificates / Performance level / Integrity level
Approvals/certificates ATEX IECEx NEPSI Show all cFMus
Protection class II 1G Ex ia IIC T6...T1 Ex ia IIC T6...T1
Typical applications
Typical applications Fertilizer Neutralization processes in chemistry Plastic production Show all Reverse osmosis Neutralization of industrial wastewater Monitoring of biological wastewater
Signal inputs/outputs and interfaces
Signal output HART® protocol 4…20 mA
More functions and options
More functions/options Error signal according to NAMUR NE 43